148-HD500 HIGH TEMPERATURE HEAT DISTORTION TESTER
22 ตุลาคม 2567
ผู้ชม 18 ผู้ชม
- This tester adapts the air circulating heating system to test the heat resistance of plastic (usually super engineering plastic) up to 500 °C.
- The specimen racks are made by glass quartz to prevent measuring disturbances due to the deflections of the racks itself.
JIS K7191-1、K7206、ASTM D648、D1525、ISO 75-1,306
Specifications are subject to change without notice. |
|
Stations |
3 Stations |
Temperature Range |
Max. 500°C (Air Chamber, Nitrogen Gas Filling Device Attached) |
Heating Speed |
120 ± 10°C/hr, 50 ± 5°C/hr, VICAT : 120 ± 10 ℃/hr, 50 ± 5°C/hr |
Bending Stress |
JIS・ISO: 1.80MPa (Method A), 0.45MPa (Method B), 8.00Mpa (Method C) |
Weight Load |
DTUL: Choose 2 weights from initial load 76.5gf to max. 3,210gf |
Displacement Measurement |
Differential Transformer: Scale 1/1,000mm, Stroke 4mm |
Pressure Foot |
DTUL: R3.0 ± 0.2mm |
Support Length |
64 ± 1mm, 100 ± 1mm |
Refrigerating Device |
Fan type (Air Fan Cooling), 3 Fans |
Oil Circulation Device |
Propeller type, 2 propellers |
Software |
Windows 10 Compatible (Included) |
Accessories |
Pressure Foot Adjustment, Specimen Holder |
Options |
Simultaneous Loading Device, Built-in Printer, Carbon Gas Injection Cooling System, Glass Sample for Calibration (Fused Quartz), Signal Tower Light |
Power Source |
AC 200V, Single-Phase, 30A, 50/60Hz |
Dimensions & Weight (Approx.) |
W1,100 × D750 × H1,560mm/ 260kg |