258-L IMPACT TESTER (WITH REFRIGERATING MACHINE)
24 ตุลาคม 2567
ผู้ชม 20 ผู้ชม
- This tester enables the operator to conduct the Charpy and Izod impact tests in a low temperature condition.
- The low temperature chamber is double-doored and the inner door is equipped with working gloves so that the temperature inside the chamber is kept constant while the operator is attaching the test specimen.
JIS K7110、K7111-1、ASTM D256、D6110、ISO 179-1:2010、180
Specifications are subject to change without notice. |
|
Temperature Range |
-35 °C to 60 °C (Refrigerator) |
Safety Device |
Hammer Setting Sensor (Interlock System on the Upper Door), Acrylic Safety Cover |
Additional Specifications |
Refer to the below table |
Option |
Software, Electric Hammer Release Device, Torque Adjustment Device (Only for Izod) |
Power Source |
AC 200 V, 3-Phase, 40 A, 50/60 Hz |
Dimensions/ Weight (Approx.) |
W1,100 × D1,160 × H1,760 mm, 450 kg |
Specifications are subject to change without notice. |
||
Test method |
Charpy Impact Tester |
Izod Impact Tester |
Reference standards |
JIS K7111, ISO 179, ASTM D6110 |
JIS K7110, ISO 180, ASTM D256 |
Capacity |
0.5, 1, 2, 4, 5, 7.5, 15, 25 J |
1, 2.75, 5, 11, 22 J |
Lift-up angle |
150° |
150° |
Impact speed (m/s)* |
2.9, 3.8 (3.46) |
3.5 (3.46) |
Striking Edge Angle* |
30 ± 1° (45 ± 2°) |
75° (-) |
Striking Edge Radius (mm)* |
R2 ± 0.5 (R3.17 ± 0.12) |
R0.8 (R0.79 ± 0.12) |
Anvil-Edge Distance (mm)* |
– |
22 ± 0.2 (22 ± 0.05) |
Support Length (mm)* |
62 + 0.5 (101.6 ± 0.5) |
– |
Accessories |
Specimen setting gauge, Block gauge, Torque wrench (Izod only) |
Specimen setting gauge, Block gauge, Torque wrench (Izod only) |