Geer Aging Oven No.102| No.103 | YASUDA SEIKI

Thermal aging tester for rubber, geer oven , plastics, and wire insulation—utilizing a forced-air circulation(gear-driven) aging oven—that controls temperature, air exchange rate (3–60 air changes per hour), and air velocity. The unit features a rotating specimen platform (operating at 7–10 rpm) and is designed to analyze the durability and stability of polymeric materials under long-term accelerated thermal aging with circulating air, enabling the prediction of the material's actual service life.

Key Feature

  • Accelerated Geer aging test oven: The specimen-rotating platform and the internal air circulation system ensure that all specimens are exposed to uniform heat.

  • Capable of measuring and controlling environmental conditions: Allows for the measurement and adjustment of air exchange rates, with multiple configurable ranges 3-200times/hr. 3 to 200 time/hrs.

  • Choose between 2 fan speed settings: low speed (0.5 ± 0.1 m/s) and high speed  (1.0 ± 0.2 m/s) Via a touchscreen, in accordance with the ISO 188 standard

  • Accommodates a wide range of workpiece sizes: available in both standard and extra-large (outsize) versions.

  • Elevate product standards: Ensure that rubber, plastic, and electrical wire materials meet the quality certification requirements of ASTM, JIS, and UL with confidence.

 

Reference Standards

JIS A5756、B7757、C3005、K6257、K6723、K7212、K7368、ASTM D573、D5374、D5423、E145、UL-746B、IEC-60216-4-1、ISO 188、4577

Specification

Specification SHF-S SHF-SA
Model Characteristics Equipped with air exchange rate measurement/display via touch panel.
Damper: Opened/closed manually.
Equipped with air exchange rate measurement/display via touch panel.
Damper: Automatic Damper Control.
Inner dimensions W450 × D450 × H500 mm
Entire Heated Space Volume 225,225 cm³
Temperature range Room Temperature + 20 to 300 °C
Air velocity Low: 0.5 ± 0.1 m/s, High: 1.0 ± 0.2 m/s (Controlled via touch panel)
Air Exchange Rate Measurement Power Consumption Method
Air Exchange Rate Control Adjustable between 3 to 10 times/hour, 5 to 20 times/hour, or over 60 times/hour
Hanger Rotation Speed 7.5 ± 2.5 rpm
Safety Device Overheat Detection Device, Circuit Breaker
Accessories Specimen Hanger (2 stages), Clip (40 pcs), Shelf (2 pcs)
Option CSV Output, Analog Voltage Output, UL Type Air Exchange Rate 100 to 200 times/hr modification
Power AC 200 V, 3-Phase, 30 A, 50/60 Hz
*When configured with UL Type Air Exchange Rate (100 to 200 times/hr): AC 200 V, 3-Phase, 40 A, 50/60 Hz
Dimensions / Weight (Approx.) W1,180 × D700 × H1,560 mm / 300 kg W1,250 × D700 × H1,650 mm / 305 kg
Reference Standards JIS: A5756, B7757, C3005, K6257, K6723, K7212, K7368
ASTM: D573, D5374, D5423, E145
UL: 746B  |  IEC: 60216-4-1  |  ISO: 188, 4577

FAQ Section

Q: How does the Geer Aging Oven (Model No. 102) differ from a standard heating oven?

A: Oven No.102 features a system for measuring and controlling the air exchange rate, as well as precisely regulating airflow direction and velocity. It is also equipped with a rotating specimen platform, ensuring that every item receives uniform heat exposure from all sides—capabilities that standard hot-air ovens lack, making them unable to maintain the specific conditions required for aging test standards.

Q: What is the difference between the SHF-S and SHF-SA models of the No.102?

A: The SHF-S model displays the air exchange rate on its screen and requires manual operation of the air damper, whereas the SHF-SA model features an automatic air damper system that maintains and regulates the air exchange rate to meet specified levels without the need for manual adjustment.

Q: Does the No.102 oven model support testing in accordance with the UL-746B standard?

A: Yes, it is supported. The unit offers an optional configuration specifically designed for the UL 746B (2014) standard, capable of achieving an air exchange rate of up to 100–200 times per hour. It also includes a data export function, allowing information to be saved as a CSV file for use as a formal reference in international test reports.

 

Geer Type Aging Oven 1

 

The oven features controlled internal air circulation, adjustable to a specified number of cycles per hour, an exhaust fan, an air flow controller (for both intake and exhaust), a temperature controller, and a fixture for holding test specimens.

 

Setting Test Conditions According to Standards 

Setting Test Conditions According to Standards

 

Accelerated Aging Test Apparatus for Predicting Material Lifespan  Geer Aging Oven No. 102-SHF-S/SA

 

Accelerated Aging Tester for Predicting Material Lifespan Geer Aging Oven No. 102 SHF SSA

 

 

Yasuda Accelerated Aging Testers come in two models: adjustable air vent in/out.Manual and automatic operation via touch screen

Both modelsThe device will automatically calculate the air exchange rate via the touch screen.

Adjust the air intake and outlet.

 

The device is equipped with a room air temperature sensor, which is an important value for calculating the air exchange rate. It transmits the measured temperature for calculating the air exchange rate. Real-time Can control the maximum temperature. 300 Degree Celsius

The device will automatically calculate the air exchange rate.

 

 

For the automatic model, if the room temperature changes, the device will automatically adjust the air vents to maintain the air exchange rate, ensuring that the testing conditions are controlled in accordance with the specified standards.

Graph of the Relationship Between Air Exchange Rate and Temperature

For your convenience, you can record your test history, including test results and temperatures., Air exchange rate, The room temperature is displayed on the touchscreen for easy use.

Graph of the Relationship Between Air Exchange Rate and Temperature

 