A fully automated melt flow index tester—the top-tier model in the melt flow testing series—utilizes a robotic mechanism to enable continuous, unattended testing of multiple samples. It can process up to 50 samples in a single run; the system offers a choice of three magazine sizes (accommodating 12, 24, or 50 samples), facilitating overnight operation. The instrument is designed to measure the Melt Flow Rate (MFR) of molten plastics and also supports the determination of the Melt Volume Rate (MVR).In accordance with test methods defined by JIS, ISO 1133-1, and ASTM D1238 standards, this solution is ideal for plastics manufacturing plants, resin producers, and quality control laboratories looking to enhance testing efficiency through automation.
Reduce the risk of human error with an automated robotic arm system. Seamlessly controls the entire process—from sample loading and testing to the cleaning of the barrel and die.
Boost your confidence with real-time MVR analysis charts. Detect air bubbles within the plastic material during testing, in real time, via computer software.
Supports continuous testing of up to 50 samples. Helps save time and increase operational efficiency in laboratories with high testing volumes.
Save on maintenance costs Designed to use only gauze as the primary consumable material for cleaning, with a durable construction built to last for over 20 years.
| Specification | Details |
|---|---|
| Reference Standards | JIS K 6719-1/2, K 6921-2, K 6922-2, K 6923-1, K 6924-1, K 6926-2, K 7210-1, ISO 1133-1, ASTM D1238 |
| Number of Test Samples | 12, 24, or 50 cups of samples |
| Die | Diameter ø2.095 ± 0.005 mm, length 8.000 ± 0.025 mm |
| Piston Head | Diameter ø9.474 ± 0.007 mm, length 6.35 ± 0.10 mm |
| Cylinder | Inner diameter ø9.550 ± 0.007 mm, length 160 mm |
| Temperature | Max. 300°C (Option: 400°C) |
| Initial Load | 0.325 kgf |
| Test Load | 1.20 – 21.60 kgf |
| Test Method | • Procedure A: manual sample cutting (Option: automatic sample cutter) • Procedure B: automatic |
| Operation & Material Handling | • Cylinder Charging: Automatic • Loading: Automatic • Weight Replacement: Manual (Option: automatic) |
| Cleaning Systems | • Cylinder Cleaning: Automatic (solvent can also be used in the automatic cleaning cycle) • Die Cleaning: Automatic • Piston Cleaning: Automatic • Cleaning Rod Cleaning: Automatic • Gauze Supply and Removal: Automatic |
| Data Processing | • Procedure A: manual (Option: automatic) • Procedure B: automatic |
| Procedure B Piston Travel Measurement | Rotary encoder |
| PC Software | Available for Windows |
| Accessories | Cleaning rods (for die and cylinder), cup for cleaning solvent, sample cup for powder samples, sample cup holder, funnel, go/no go gauge, wastebasket, gauze, air blow gun |
| Power Source | AC 200V, 1-phase, 15 A, 50/60 Hz (20 A if installed with sample drying oven) |
| Air Source | 0.5 MPa or above |
| Dimensions and Weight (Approx.) | • 12 or 24 samples: W950 × D680 × H1,720 mm, 350 kg • 50 samples: W1,100 × D700 × H1,750 mm, 400 kg |
A: It supports a wide range of forms—such as pellets, granules, sheets, and powders—by utilizing a specialized sample cup with a Teflon-coated interior, which ensures smooth, jam-free feeding of the sample into the cylinder.
A: At the end of the test, the robotic arm removes any excess plastic residue and uses prepared gauze and cleaning rods to thoroughly wipe the inner walls of the cylinder and the die head, ensuring that subsequent tests are highly accurate and free from contamination.
A: This can be done by choosing to install optional features. Automatic Weight Loading Device The system automatically adjusts the load weight based on parameters set within the computer software, helping to reduce the risk of injury associated with manually lifting heavy loads.