HIGH TEMPERATURE HEAT DISTORTION TESTER NO.148-HD-500 | YASUDA SEIKI

High-temperature heat deflection temperature (HDT) tester utilizes an air-circulating heating system to evaluate the heat resistance of plastics at temperatures up to 500 °C (typically used for advanced engineering plastics or "super engineering plastics"). The specimen racks are constructed from quartz glass to prevent measurement inaccuracies that could otherwise arise from the thermal expansion or deformation of the racks themselves when exposed to high temperatures.

Key Feature

  • TEST Capable of controlling heat and measuring the compression/torsion of 3 test specimens.

  • High measurement precision and stability: Use a high-resolution LVDT sensor 0.001 mm Measure the elongation and flexure of plastic specimens in accordance with international standards.

  • Test temperature High heat 500°C (No.148-HD-500)

  • Rapid analysis of results via software: Control the heating process and record the deformation-versus-temperature graph using a Windows-based system.

Reference Standards

JIS K7191-1、K7206、ASTM D648、D1525、ISO 75-1,306

Specification

Model 148-HD500
Stations 3 stations
Temperature Max. 500°C (Air chamber, nitrogen purging unit)
Heating Rate 120 ± 10°C/hr, 50 ± 5°C/hr
Cooling System Air cooling by propeller fans (2 fans)
Flexural Stress • JIS / ISO: 1.80 MPa, 0.45 MPa, 8.0 MPa
• ASTM: 1.82 MPa, 0.455 MPa
Test Load • DTUL: select 2 loads from 76.5 gf (initial load) to max. 3,210 gf
• VST (Option): 10 ± 0.2 N, 50 ± 1 N
• BP (Option): 20–0.4 N
Displacement Measurement Sensor type: LVDT (resolution 1/1,000 mm, stroke length 0–4 mm)
Loading Edge • DTUL: radius R3.0 ± 0.2 mm
• VST (Option): indenting tip 1.000 ± 0.015 mm²
• BP (Option): steel ball diameter ø5 mm
Span Between Supports 64 ± 1 mm, 100 ± 2 mm
PC Software Available for Windows
Accessories Loading edge setting gauge, sample holders
Options Safety cover, sample dummy for distortion check, CO2 cooling system, simultaneous loading system (selectable only for VST specifications)
Power Source AC 200 V, 1-phase, 30 A, 50/60 Hz
Dimensions and Weight (Approx.) W1,100 × D750 × H1,560 mm, 260 kg

FAQ Section

Q: What are the differences between HDT (DTUL) and Vicat (VST) testing?

A: HDT (Heat Deflection Temperature) is the measurement of the temperature at which a plastic rod bends to a specified distance under bending stress. Vicat (VST) involves measuring the temperature at which a flat-tipped probe with an area of ​​1 mm² is pressed 1 mm deep into the plastic material. No.148 Series can support both types of testing by changing the indenting tip.

Q: Which model should be chosen for testing plastics that can withstand temperatures above 300°C?

A: Model selection guide No.148-HD-500 (High Temperature Type) utilizes an air circulation heating system and a quartz glass specimen-holding structure, enabling high-temperature testing up to 500°C without the structure itself expanding and affecting the measurement values

Q: Does the No.148 testing machine feature a system to help lower the oil bath temperature after a test is completed?

A: มี The unit comes equipped with water cooling coils for circulating water to lower the temperature of the oil bath, and it also supports the installation of optional accessories. Built-in Refrigeration Unit or a cooling system, to rapidly lower the oil bath temperature back to 20–23°C, making it immediately ready for the next test run.