เครื่องทดสอบอ่อนตัวของพลาสติกอัตโนมัติ HDT / Vicat No.148-HDA (Yasuda Seiki)

The No.148-HDA Automatic HDT/Vicat Tester from Yasuda Seiki is a fully automated system designed to analyze the heat resistance properties of plastics and polymers. It measures the Deflection Temperature Under Load (DTUL) and the Vicat Softening Temperature (VST) in compliance with international standards, including ASTM D648, ASTM D1525, ISO 75, ISO 306, and JIS K7191. This instrument is ideal for quality control (QC) laboratories and research and development (R&D) centers within the plastic resin and automotive components industries that require high precision and the capacity for high-volume, continuous testing. Simply place the specimens into the specimen feeder, and the computer software will direct the instrument to perform heat resistance testing on up to 120 samples in a continuous sequence.

Key Feature

  • Reduce the workload on operators with an automated robotic sample-feeding system: capable of continuously feeding and testing up to 120 samples, thereby saving time and minimizing human error during testing.

  • Comprehensive support for international standards in a single unit: capable of accurately performing HDT (DTUL), Vicat Softening Temperature (VST), and Ball Pressure (BP) tests.

  • Temperature control and micron-level displacement measurement: Capable of reaching temperatures up to 300°C, the system incorporates an LVDT deformation sensor with a resolution of 1/1,000 mm, ensuring maximum stability in analysis results.

  • Intelligent control and data logging software: Operates on a Windows-based PC, allowing for the configuration of test conditions, data recording, graphical visualization, and the generation of test summary reports.

Reference Standards

JIS K7191-1、K7206、ASTM D648、D1525、ISO 75-1、306

Specification

Specification Item Technical Details
Stations 3 or 6 Stations
Temperature Range Max. 300°C (Oil Bath)
Heating Speed 120 ± 10°C/hr and 50 ± 5 °C/hr, VICAT: 120 ± 10℃/hr, 50 ± 5°C/hr
Bending Stress JIS・ISO: 1.80MPa (Method A), 0.45MPa (Method B), 8.00Mpa (Method C)
ASTM: 1.82Mpa, 0.455Mpa
Weight Load DTUL: Choose 2 weights from initial load 76.5gf to max. 3,210gf
*Simultaneous Loading Device included
Options: VICAT: 10 ± 0.2N, 50 ± 1N / BP: 19.6 ± 20N
Displacement Measurement Differential Transformer: Scale 1/1,000 mm, Stroke 4 mm
Pressure Foot DTUL: R3.0 ± 0.2mm
Options: VICAT: 1.000 ± 0.015mm² / BP: Ball Diameter 5mm
Support Length 64 ± 1mm, 100 ± 1mm (Option: 101.6 ± 0.5mm)
Refrigerating Device Water Circulating System, Built-In Refrigerator
*For low test start temperature (e.g., 20 to 23°C)
Oil Circulation Device Propeller Type
For 3 and 4 Stations: 2 Propellers
For 6 Stations: 3 Propellers
Test Specimens 40 pcs × 3 sets (Total 120 pcs)
*For T10mm
Software Windows 10 compatible (Included)
Accessories Pressure Foot Adjustment, Specimen Holder
Options Silicon Oil, Glass sample for calibration (fused quartz), Built-in printer, Oil level sensor, Safety door lock, Signal tower light
Power Source
Without refrigerator:
• For 3 and 4 Stations: Single-Phase, AC200/220V 20A 50/60Hz
• For 6 Stations: Single-Phase, AC200/220V 30A 50/60Hz
With refrigerator:
• For 3 and 4 Stations: Three-Phase, AC200/220V 20A 50/60Hz
• For 6 Stations: Three-Phase, AC200/220V 30A 50/60Hz
Air Source 0.5MPa or more
Dimensions & Weight (Approx.) W1,460 × D900 × H1,715mm
450 to 550kg

FAQ Section

Q1: How many specimens can the HDT/Vicat Tester (Model No. 148-HDA) accommodate for sequential testing?

A: The system features a robotic arm and cartridges for holding test specimens. It can accommodate up to 390 specimens (assuming a thickness of 4 mm) and automatically feed them for continuous testing—handling as many as 120 specimens per cycle—without requiring the machine to stop.

Q2: What types of tests can be performed with model No.148-HDA?

A: It supports two primary types of tests: Deflection Temperature Under Load (DTUL/HDT) and Vicat Softening Temperature (VST). Additionally, a Ball Pressure (BP) test head can be installed as an optional feature.

Q3: How does the oil bath cooling system function to prepare for the next test?

A: The system is equipped with cooling coils integrated with an industrial refrigeration unit, allowing the oil temperature to be rapidly lowered back to the initial range of 20–23°C after each test cycle. This effectively reduces waiting time and increases the number of tests that can be performed per day.